Thermogravimetric Analyser (TGA)
Weighs a sample continuously while it is heated, and records mass loss against temperature. Used to measure water and solvent content, decomposition and residue.
- Pharmaceutical
- Polymers
- Materials science
- Minerals
- Chemicals

TGA calibration procedures
Temperature calibration (Curie point)
Uses a magnet and ferromagnetic standards that lose their magnetism at known Curie temperatures to check the TGA temperature scale.
Open procedureMass calibration
Checks the TGA microbalance with certified milligram weights, and optionally its overall performance with calcium oxalate monohydrate.
Open procedureInstrument qualification (DQ · IQ · OQ · PQ)
- 01
DQ — Design qualification
Defines temperature range, balance sensitivity, gases, autosampler and software compliance needs.
- 02
IQ — Installation qualification
Confirms correct installation, gas and exhaust connections, software version and documentation.
- 03
OQ — Operational qualification
Checks temperature accuracy (Curie point standards), mass accuracy (certified weights) and baseline stability.
- 04
PQ — Performance qualification
Shows routine performance, for example with calcium oxalate monohydrate whose three mass-loss steps are known from its formula.
Keep every TGA in control: usage logs, calibration and maintenance schedules, and automatic blocking of overdue instruments.