Thermogravimetric Analyser (TGA)

Weighs a sample continuously while it is heated, and records mass loss against temperature. Used to measure water and solvent content, decomposition and residue.

  • Pharmaceutical
  • Polymers
  • Materials science
  • Minerals
  • Chemicals
Photo of a Thermogravimetric Analyser (TGA)
Photo: Luigi Chiesa, CC BY 3.0 (cropped), via Wikimedia Commons
CALIBRATION

TGA calibration procedures

QUALIFICATION

Instrument qualification (DQ · IQ · OQ · PQ)

  1. 01

    DQ — Design qualification

    Defines temperature range, balance sensitivity, gases, autosampler and software compliance needs.

  2. 02

    IQ — Installation qualification

    Confirms correct installation, gas and exhaust connections, software version and documentation.

  3. 03

    OQ — Operational qualification

    Checks temperature accuracy (Curie point standards), mass accuracy (certified weights) and baseline stability.

  4. 04

    PQ — Performance qualification

    Shows routine performance, for example with calcium oxalate monohydrate whose three mass-loss steps are known from its formula.