TGA · Calibration procedure
Mass calibration
Checks the TGA microbalance with certified milligram weights, and optionally its overall performance with calcium oxalate monohydrate.

PURPOSE
Why this is checked
Every TGA result is a percentage mass loss, so the balance must read mass correctly across the sample range.
Frequency
Typically every 6–12 months, with a performance check more often as defined in your SOP.
Learn how the TGA worksYOU WILL NEED
Standards & tools
- Certified milligram weights (for example 10, 50 and 100 mg) traceable to national standards
- Optional: calcium oxalate monohydrate for a performance check (theoretical steps about 12.3% water, 19.2% carbon monoxide and 30.1% carbon dioxide)
BEFORE YOU START
Prerequisites
- Balance stable and furnace at room temperature
- Weights handled with tweezers only
PROCEDURE
Step by step
- 1Tare the empty pan.
- 2Place each certified weight in the pan and record the reading once stable.
- 3Calculate the error for each weight.
- 4Optionally run calcium oxalate monohydrate from ambient to about 900 °C and compare each mass-loss step with theory.
RECORD
Example observation table
| Certified mass (mg) | Observed (mg) | Error (mg) | % error |
|---|---|---|---|
| 10.000 | 10.004 | +0.004 | +0.04 |
| 50.000 | 49.987 | −0.013 | −0.026 |
| 100.000 | 100.021 | +0.021 | +0.021 |
CALCULATION
How to calculate the result
% error = (Observed − Certified) ÷ Certified × 100
- 50 mg weight: (49.987 − 50.000) ÷ 50.000 × 100 = −0.026%
- Calcium oxalate monohydrate (M = 146.11 g/mol): water 18.02 ÷ 146.11 × 100 = 12.33%
ACCEPTANCE CRITERIA
Pass if
- Error within the limit in your SOP or the manufacturer's specification (commonly about ±0.1% or a stated µg value).
IF IT FAILS
What to do
- Label the instrument 'Out of calibration' and stop using it.
- Investigate the cause and arrange adjustment or repair.
- Assess the impact on results generated since the last passing calibration.
- Repeat the full calibration before returning it to use.