TGA · Calibration procedure

Temperature calibration (Curie point)

Uses a magnet and ferromagnetic standards that lose their magnetism at known Curie temperatures to check the TGA temperature scale.

Photo of a TGA
Photo: Luigi Chiesa, CC BY 3.0 (cropped), via Wikimedia Commons
PURPOSE

Why this is checked

The TGA thermocouple does not touch the sample, so the reported temperature can differ from the true sample temperature. Curie point standards give sharp, known reference points across the range.

Frequency

Typically every 6–12 months, and after furnace or thermocouple work, as defined in your SOP.

Learn how the TGA works
YOU WILL NEED

Standards & tools

  • Certified Curie point standards, for example nickel (about 355–358 °C) and others such as alumel or Perkalloy, as the range requires; use the certificate values
  • Permanent magnet placed outside the furnace as the manufacturer directs
BEFORE YOU START

Prerequisites

  • Same purge gas, flow and heating rate as routine methods
  • Clean pan and stable balance
PROCEDURE

Step by step

  1. 1Place a small piece of the standard in the pan and position the magnet so it pulls on the sample.
  2. 2Tare and heat through the Curie point at the routine heating rate (for example 10 °C/min).
  3. 3At the Curie point the magnetic pull disappears and the apparent mass changes suddenly.
  4. 4Determine the temperature of the step (onset or midpoint, as the certificate defines).
  5. 5Repeat for each standard and adjust the temperature calibration in the software if needed.
RECORD

Example observation table

StandardCertified (°C)Observed (°C)Difference (°C)
Nickel358.0357.2−0.8
CALCULATION

How to calculate the result

Difference = Observed Curie temperature − Certified Curie temperature
  1. Nickel: certified 358.0 °C (illustrative), observed 357.2 °C
  2. Difference = 357.2 − 358.0 = −0.8 °C
ACCEPTANCE CRITERIA

Pass if

  • Difference within the limit in your SOP (commonly about ±2 °C to ±5 °C, depending on the instrument and use).
IF IT FAILS

What to do

  • Label the instrument 'Out of calibration' and stop using it.
  • Investigate the cause and arrange adjustment or repair.
  • Assess the impact on results generated since the last passing calibration.
  • Repeat the full calibration before returning it to use.