Thermogravimetric Analyser (TGA)

Weighs a sample continuously while it is heated, and records mass loss against temperature. Used to measure water and solvent content, decomposition and residue.

  • Pharmaceutical
  • Polymers
  • Materials science
  • Minerals
  • Chemicals
Photo of a Thermogravimetric Analyser (TGA)
Photo: Luigi Chiesa, CC BY 3.0 (cropped), via Wikimedia Commons
PREVENTIVE MAINTENANCE

Maintenance checklist

Each use

  • Check gas flow and balance stability
  • Inspect the pan and hang-down wire

Monthly / as needed

  • Clean the furnace tube and pans
  • Check the thermocouple position

Periodic (as per SOP)

  • Temperature calibration with Curie point standards
  • Mass calibration with certified weights
  • Performance check with calcium oxalate monohydrate

Annual

  • Preventive maintenance by the service engineer
SAFETY

Safety precautions

Before you work on the TGA
  • The furnace may reach 1000 °C or more; let it cool before handling parts.
  • Decomposition gases can be toxic; vent the exhaust to an extraction system.
  • Secure gas cylinders and check regulators.
  • Do not test explosive or highly energetic materials without a risk assessment.