Dissolution tester calibration procedures
Mechanical calibration (shaft, vessel, centring)
Measures the physical set-up of every position: shaft wobble, shaft and vessel verticality, centring of the shaft in the vessel, and the height of the paddle or basket above the vessel bottom.
Open procedureRotation speed
Checks that every shaft turns at the set speed, using a calibrated tachometer, at the speeds your methods use.
Open procedureTemperature
Confirms the medium in every vessel is held at 37 ± 0.5 °C, using a calibrated thermometer.
Open procedurePerformance verification test
Runs a dissolution test on reference tablets supplied by the pharmacopoeia (for USP, prednisone tablets RS) and compares the results with the acceptance ranges given for that lot.
Open procedureInstrument qualification (DQ · IQ · OQ · PQ)
- 01
DQ — Design qualification
States the lab's needs (apparatus types, number of positions, speed range, temperature control, auto-sampling, data integrity of the controller) and confirms the selected model meets compendial dimensional requirements.
- 02
IQ — Installation qualification
Confirms delivery and installation: serial numbers, vessels, paddles and baskets with their certificates of dimensions, a level and vibration-free bench, utilities and documentation.
- 03
OQ — Operational qualification
Checks speed, temperature, shaft wobble, shaft and vessel verticality, centring and paddle or basket height at every position, plus timer and auto-sampler functions.
- 04
PQ — Performance qualification
Shows the system performs for its intended use, through a performance verification test with reference tablets and/or a documented enhanced mechanical calibration, as your SOP requires.
Keep every Dissolution tester in control: usage logs, calibration and maintenance schedules, and automatic blocking of overdue instruments.