LimsPulse
Dosage-form testing

Dissolution Test Apparatus

Measures how fast and how completely the drug is released from a tablet or capsule into a liquid medium held at body temperature. A key quality and release test for oral solid dosage forms.

  • Pharmaceutical
  • Generic drug development
  • Nutraceuticals and dietary supplements
  • Veterinary medicines
  • Contract research and testing labs
Photo of a Dissolution Test Apparatus
Photo: Gpkp, CC BY-SA 4.0 (cropped), via Wikimedia Commons
PREVENTIVE MAINTENANCE

Maintenance checklist

Each use

  • Check bath level and temperature
  • Inspect paddles, baskets and vessels
  • Clean vessels, shafts and baskets after use
  • Record any abnormal noise or vibration

Weekly

  • Clean the bath and change the water (use an algicide only if it does not leave residues)
  • Check vessel centring rings and plate fixings
  • Clean auto-sampler lines and replace in-line filters if used

Monthly / quarterly

  • Check basket mesh for damage or corrosion and replace worn baskets
  • Check drive belt and shaft bearings for noise
  • Verify speed and temperature with calibrated tools

Six-monthly / annual

  • Full mechanical calibration of all positions
  • Performance verification test where required by your SOP
  • Preventive maintenance by a trained engineer
SAFETY

Safety precautions

Before you work on the Dissolution tester
  • Keep hands and loose clothing away from rotating shafts; stop rotation before adjusting anything in the vessel.
  • The bath contains electrical heaters: never run it without enough water and keep the electrical parts dry.
  • Some media contain surfactants, acids or enzymes; read the safety data sheet and wear gloves and safety glasses.
  • Handle glass vessels with care; chipped vessels can cut and must be replaced.
  • Wipe up spills promptly; wet floors around the bath are a slip hazard.
In LimsPulse · Calibration Management

Keep every Dissolution tester in control: usage logs, calibration and maintenance schedules, and automatic blocking of overdue instruments.