Dissolution tester · Calibration procedure

Rotation speed

Checks that every shaft turns at the set speed, using a calibrated tachometer, at the speeds your methods use.

Photo of a Dissolution tester
Photo: Gpkp, CC BY-SA 4.0 (cropped), via Wikimedia Commons
PURPOSE

Why this is checked

Agitation speed has a direct effect on the dissolution rate. A shaft running fast or slow gives a biased result.

Frequency

Typically every 6 months and after drive maintenance; some labs also check speed before each test.

Learn how the Dissolution tester works
YOU WILL NEED

Standards & tools

  • Calibrated optical or contact tachometer, traceable to national standards
BEFORE YOU START

Prerequisites

  • Paddles or baskets fitted
  • Vessels filled with medium or water as your SOP requires
  • Apparatus warmed up
PROCEDURE

Step by step

  1. 1Set the speed to the lowest routine value (e.g. 50 rpm) and allow it to stabilise.
  2. 2Measure the speed of each shaft with the tachometer and record it.
  3. 3Repeat at the other speeds used (e.g. 75 and 100 rpm).
  4. 4Calculate the % deviation from the set value for each reading.
RECORD

Example observation table

PositionSet (rpm)Observed (rpm)% deviation
15049.9−0.2
25050.1+0.2
35050.00.0
45049.8−0.4
55050.2+0.4
65050.00.0
110099.6−0.4
2100100.3+0.3
3100100.1+0.1
410099.8−0.2
5100100.4+0.4
610099.9−0.1
CALCULATION

How to calculate the result

% deviation = (Observed − Set) ÷ Set × 100
  1. Set 50 rpm, observed 49.8 rpm at position 4
  2. % deviation = (49.8 − 50) ÷ 50 × 100 = −0.4%
  3. Within ±4% (±2 rpm at 50 rpm)
ACCEPTANCE CRITERIA

Pass if

  • USP: within ±4% of the stated speed.
  • Enhanced mechanical calibration practice often uses a tighter limit, such as ±2% or ±2 rpm; follow your SOP.
IF IT FAILS

What to do

  • Stop using the apparatus and label it.
  • Check the drive belt, motor and speed controller; get it serviced.
  • Recalibrate and assess the impact on earlier results.