A dry heating block that holds sealed glass tubes at 150 °C for two hours so that potassium dichromate can oxidise the organic matter in water and wastewater samples. It is the heart of the closed reflux COD test.

  • Environmental testing laboratories
  • Municipal water and wastewater utilities
  • Chemical and pharmaceutical effluent treatment
  • Food and beverage processing
  • Textile, pulp and paper, and distilleries
Illustration of COD Digestion Reactor
CALIBRATION

COD digester calibration procedures

QUALIFICATION

Instrument qualification (DQ · IQ · OQ · PQ)

  1. 01

    Design qualification (DQ)

    Define the tube size, number of wells, temperature range (at least 150 °C, with options for 100–120 °C digestions) and timer features needed for the lab's COD and nutrient tests.

  2. 02

    Installation qualification (IQ)

    Record model and serial numbers, install on a level, heat-resistant bench away from flammables, and check the power supply and safety cover.

  3. 03

    Operational qualification (OQ)

    Verify block temperature at 150 °C in several wells with a calibrated thermometer, and verify the timer. Typical limits are 150 ± 2 °C.

  4. 04

    Performance qualification (PQ)

    Digest replicate KHP standards and real samples and show recoveries and repeatability meet the method's quality control limits.