Autoclave (steam steriliser)

Kills microorganisms, including bacterial spores, with saturated steam under pressure. Used in the QC microbiology lab to sterilise culture media, glassware and equipment, and to decontaminate used media before disposal.

  • Pharmaceutical
  • Biotech
  • Hospitals
  • Food & beverage
  • Research laboratories
  • Medical devices
Photo of a Autoclave (steam steriliser)
Photo: Nadina Wiórkiewicz (Nadine90), CC BY-SA 3.0 (cropped), via Wikimedia Commons
CALIBRATION

Autoclave calibration procedures

QUALIFICATION

Instrument qualification (DQ · IQ · OQ · PQ)

  1. 01

    DQ — Design qualification

    Defines the needs: chamber size, cycles required (liquid, porous, decontamination), air removal type, steam source, data recording with audit trail, and safety features.

  2. 02

    IQ — Installation qualification

    Confirms the autoclave, utilities (steam, water, drain, power) and documentation are installed as specified, with calibrated sensors and the correct software version.

  3. 03

    OQ — Operational qualification

    Tests the empty chamber: alarms and interlocks, vacuum leak test, Bowie-Dick test, and empty-chamber temperature distribution with thermocouples.

  4. 04

    PQ — Performance qualification

    Proves each load pattern is sterilised: heat penetration with thermocouples in the coldest containers, F0 calculation and biological indicator challenge, usually in three consecutive runs.