LimsPulse
Chromatography

Ultra-High Performance Liquid Chromatography (UHPLC)

A high-pressure form of HPLC that uses columns packed with very small particles to give faster runs, sharper peaks and less solvent use. Used in pharma QC for fast assays, impurity profiling and high-throughput testing.

  • Pharmaceutical
  • Biotech
  • Contract research (CRO / CDMO)
  • Food & beverage
  • Environmental
  • Clinical research
Photo of a Ultra-High Performance Liquid Chromatography (UHPLC)
Photo: Groep29 BFW, Public domain (cropped), via Wikimedia Commons
CALIBRATION

UHPLC calibration procedures

QUALIFICATION

Instrument qualification (DQ · IQ · OQ · PQ)

  1. 01

    DQ — Design qualification

    States the user requirements: pressure rating, dwell volume, flow range, injection range, detector type and cell volume, data rate and a compliant CDS.

  2. 02

    IQ — Installation qualification

    Confirms delivery, installation, utilities, firmware and software versions, and that the system is connected to the CDS correctly.

  3. 03

    OQ — Operational qualification

    Tests flow accuracy, gradient accuracy, injector precision, carry-over, detector wavelength accuracy, noise and linearity, and column temperature against specifications.

  4. 04

    PQ — Performance qualification

    Shows the system performs consistently for its intended methods, for example with routine system suitability and a periodic holistic test mixture.

In LimsPulse · Calibration Management

Keep every UHPLC in control: usage logs, calibration and maintenance schedules, and automatic blocking of overdue instruments.

See how