Checks that the UHPLC autosampler injects the same small volume every time, using replicate injections of a caffeine solution.

Photo of a UHPLC
Photo: Groep29 BFW, Public domain (cropped), via Wikimedia Commons
PURPOSE

Why this is checked

UHPLC injections are small (often 1–2 µL), so any air bubble, worn seal or partial blockage gives a large relative error. Injector precision underpins every quantitative result and system suitability %RSD.

Frequency

Typically every 6 months, and after major repair or relocation, as defined in your calibration SOP.

Learn how the UHPLC works
YOU WILL NEED

Standards & tools

  • Caffeine reference solution (e.g. about 20–50 µg/mL in water)
  • Short UHPLC C18 column (e.g. 50 × 2.1 mm, sub-2 µm)
  • HPLC-grade water and methanol
BEFORE YOU START

Prerequisites

  • Equilibrate with a suitable mobile phase (e.g. water : methanol 80 : 20) until the baseline is stable
  • Set the detector to 273 nm and a data rate suited to the narrow peak
  • Prime the needle wash and make sure the sample syringe is free of air
PROCEDURE

Step by step

  1. 1Inject the caffeine solution six times at the routine volume (e.g. 1 µL).
  2. 2Integrate each peak with the same processing method.
  3. 3Calculate the mean, standard deviation and %RSD of the peak areas.
  4. 4If your SOP requires, repeat at a second volume (e.g. 5 µL) and inject a blank after the highest standard to check carry-over.
RECORD

Example observation table

InjectionVolume (µL)Peak area
11.0285,412
21.0286,031
31.0284,977
41.0285,760
51.0286,288
61.0285,104
CALCULATION

How to calculate the result

%RSD = (Standard deviation ÷ Mean) × 100
  1. Mean area = 285,595
  2. Standard deviation = 520
  3. %RSD = 520 ÷ 285,595 × 100 = 0.18%
Use the Mean, SD and %RSD calculator
ACCEPTANCE CRITERIA

Pass if

  • %RSD of peak area typically not more than 1.0% (n = 6) at the routine volume; follow the manufacturer's specification for very small volumes and your SOP.
IF IT FAILS

What to do

  • Check the syringe, needle, needle seat and rotor seal for wear or blockage.
  • Purge the syringe and wash lines to remove air.
  • Service the autosampler and repeat the test.
  • Assess the impact on results since the last passing calibration.