LimsPulse
UHPLC · Calibration procedure

Pump flow rate accuracy

Checks that the UHPLC pump delivers the set flow by weighing the water collected over a timed interval.

Photo of a UHPLC
Photo: Groep29 BFW, Public domain (cropped), via Wikimedia Commons
PURPOSE

Why this is checked

UHPLC methods often run at low flows (0.2–0.6 mL/min) where a small absolute error is a large percentage error. Flow errors shift retention times and change peak areas, so flow is verified gravimetrically across the working range.

Frequency

Typically every 6 months, and after major repair or relocation, as defined in your calibration SOP.

Learn how the UHPLC works
YOU WILL NEED

Standards & tools

  • Calibrated analytical balance (0.1 mg readability)
  • Calibrated stopwatch or the CDS timer
  • Calibrated thermometer
  • HPLC-grade water
  • Tared, capped collection vial
BEFORE YOU START

Prerequisites

  • Replace the column with a restriction capillary that gives a back-pressure of roughly 100–200 bar (or as the manufacturer specifies)
  • Purge the pump and check that there are no leaks
  • Allow the flow to stabilise for at least 5 minutes at each setting
  • Record the water temperature to choose the correct density
PROCEDURE

Step by step

  1. 1Set 100% water at 0.2 mL/min.
  2. 2Place the outlet into the tared vial and collect for exactly 10.00 minutes; cap the vial to prevent evaporation.
  3. 3Weigh the vial and record the net mass of water.
  4. 4Convert mass to volume using the density of water at the measured temperature (e.g. 0.9970 g/mL at 25 °C).
  5. 5Repeat at 0.5 mL/min and 1.0 mL/min, or at the flows your methods use.
RECORD

Example observation table

Set flow (mL/min)Time (min)Mass (g)Volume (mL)Actual flow (mL/min)% error
0.210.001.9911.9970.1997−0.15
0.510.004.9824.9970.4997−0.06
1.05.004.9975.0121.0024+0.24
CALCULATION

How to calculate the result

Volume = Mass ÷ Density; Actual flow = Volume ÷ Time; % error = (Actual − Set) ÷ Set × 100
  1. Set 0.2 mL/min; 1.991 g collected in 10.00 min at 25 °C (density 0.9970 g/mL)
  2. Volume = 1.991 ÷ 0.9970 = 1.997 mL
  3. Actual flow = 1.997 ÷ 10.00 = 0.1997 mL/min
  4. % error = (0.1997 − 0.2) ÷ 0.2 × 100 = −0.15%
Use the Flow rate accuracy calculator
ACCEPTANCE CRITERIA

Pass if

  • Actual flow typically within ±2.0% of the set value at each flow (some SOPs use ±1.0%, or the manufacturer's specification).
IF IT FAILS

What to do

  • Label the instrument 'Out of calibration' and stop using it.
  • Check for leaks, worn pump seals and faulty check valves; arrange service.
  • Assess the impact on results generated since the last passing calibration.
  • Repeat the calibration after repair.
MORE

Other UHPLC calibrations

In LimsPulse · Calibration Management

LimsPulse schedules this calibration, records the result against limits, and blocks the instrument everywhere if it fails or goes overdue.

See how