PREVENTIVE MAINTENANCE

Maintenance checklist

Daily / each use

  • Record system pressure at the start of the run
  • Check needle-wash and seal-wash levels
  • Check for leaks at the column fittings
  • Replace mobile phases rather than topping up

Weekly

  • Replace in-line filter elements if pressure has risen
  • Clean solvent bottles and inlet filters
  • Check the autosampler needle and seal for deposits

Monthly / quarterly

  • Check lamp energy and hours
  • Review pressure ripple and replace check valves or seals if it is rising
  • Clean the flow cell if baseline noise increases

Six-monthly / annual

  • Preventive maintenance by a trained engineer
  • Calibration of flow, injector precision and detector wavelength
SAFETY

Safety precautions

Before you work on the UHPLC
  • Solvent jets at UHPLC pressure can penetrate skin: never loosen fittings or touch leaks while the system is pressurised.
  • Organic solvents are flammable and toxic; prepare mobile phases in a fume hood.
  • Do not look directly at the UV lamp.
  • Collect waste in closed, labelled containers and dispose of it as per site procedure.
  • Wear safety glasses, gloves and a lab coat.