UHPLC · Calibration procedure

Detector wavelength accuracy

Confirms the UHPLC UV / PDA detector reads the correct wavelength, using the built-in holmium oxide filter or a caffeine solution.

Photo of a UHPLC
Photo: Groep29 BFW, Public domain (cropped), via Wikimedia Commons
PURPOSE

Why this is checked

A wavelength error reduces sensitivity and changes peak areas, especially for methods that measure on the slope of an absorption band. PDA peak-purity checks also depend on an accurate wavelength scale.

Frequency

Typically every 6 months, and after major repair or relocation, as defined in your calibration SOP.

Learn how the UHPLC works
YOU WILL NEED

Standards & tools

  • Detector's built-in holmium oxide filter (bands near 361, 418 and 536 nm; use the reference values from the filter certificate or manufacturer)
  • or caffeine solution in water (maxima near 205 and 273 nm)
BEFORE YOU START

Prerequisites

  • Warm up the lamp fully (often 30–60 minutes)
  • Flush the flow cell with water and make sure it is free of bubbles
PROCEDURE

Step by step

  1. 1Run the detector's built-in wavelength verification with the holmium oxide filter, or fill the cell with caffeine solution.
  2. 2Record the wavelength of maximum absorbance for each band.
  3. 3Compare each observed maximum with the reference value.
  4. 4Print or save the detector's verification report with the raw data.
RECORD

Example observation table

Reference (nm)Observed (nm)Difference (nm)
360.8 (Ho)361.1+0.3
418.5 (Ho)418.2−0.3
536.4 (Ho)536.9+0.5
273 (caffeine)2730
CALCULATION

How to calculate the result

Difference = Observed maximum − Reference maximum
  1. Reference 536.4 nm, observed 536.9 nm
  2. Difference = 536.9 − 536.4 = +0.5 nm (within ±1 nm)
ACCEPTANCE CRITERIA

Pass if

  • Each observed maximum typically within ±1 nm (PDA) or ±2 nm (variable-wavelength detector) of the reference value, as per your SOP and the manufacturer's specification.
IF IT FAILS

What to do

  • Check lamp energy and hours; replace the lamp if low.
  • Run the manufacturer's wavelength recalibration and repeat the check.
  • Label the detector out of calibration until it passes.