Texture analyser calibration procedures

QUALIFICATION

Instrument qualification (DQ · IQ · OQ · PQ)

  1. 01

    DQ — Design qualification

    Defines force range and load cells, distance and speed range, probes and fixtures, temperature control and software requirements.

  2. 02

    IQ — Installation qualification

    Confirms the frame, load cells, accessories and software are installed on a stable, level bench as specified, with serial numbers and certificates recorded.

  3. 03

    OQ — Operational qualification

    Verifies force accuracy with traceable weights, distance accuracy with gauge blocks, speed accuracy and overload protection.

  4. 04

    PQ — Performance qualification

    Shows the system gives repeatable results on a reference material or control product under routine conditions.