LimsPulse
Texture analyser · Calibration procedure

Force and distance calibration

Verifies the load cell against traceable weights across its range and the probe travel against calibrated gauge blocks, so firmness and distance-based results can be trusted.

Photo of a Texture analyser
Photo: Altair666, CC BY-SA 4.0 (cropped), via Wikimedia Commons
PURPOSE

Why this is checked

Texture results are forces and areas under force–distance curves. A load cell that reads 2% high, or an arm that travels 0.2 mm further than set, shifts every firmness and work value. Daily calibration with the instrument's own weight sets the zero and span; a periodic traceable check across the range proves accuracy and linearity.

Frequency

Force calibration with the supplied certified weight daily or before each session; full traceable force and distance verification typically every 6–12 months and after an overload or repair, as defined in your SOP.

Learn how the Texture analyser works
YOU WILL NEED

Standards & tools

  • Traceable calibrated weights (for example 500 g, 1 kg, 2 kg and 5 kg for a 5 kg load cell; OIML class M1 or better)
  • Calibrated gauge blocks or a height gauge (for example 5 mm and 10 mm)
  • Stopwatch or the instrument's time log for speed checks
BEFORE YOU START

Prerequisites

  • Instrument level, on a stable bench and warmed up as the manual says
  • Correct load cell fitted and recognised by the software
  • Weight platform or calibration pan clean
PROCEDURE

Step by step

  1. 1Run the daily force calibration routine with the supplied weight (zero, then span).
  2. 2Place each traceable weight in turn on the calibration platform and record the force reading in newtons (or grams force).
  3. 3Calculate the expected force: F = m × g, with g = 9.80665 m/s² (or the local value if the SOP requires).
  4. 4Calculate the error for each weight as a percentage of the expected force.
  5. 5Run the height calibration so that the probe zero is at the platform surface.
  6. 6Place a gauge block under a flat probe, move the probe down until the trigger force is reached and record the height displayed; repeat with a second block.
  7. 7Optionally time a long travel at a set speed to check speed accuracy.
  8. 8Record results, weight and gauge certificates and the load cell serial number.
RECORD

Example observation table

AppliedExpectedReadingErrorResult
500 g4.903 N4.893 N−0.21%Pass
1000 g9.807 N9.818 N+0.12%Pass
2000 g19.613 N19.580 N−0.17%Pass
5000 g49.033 N49.120 N+0.18%Pass
Gauge block 10.00 mm10.00 mm10.02 mm+0.02 mmPass
CALCULATION

How to calculate the result

Formula
Expected force F = m × 9.80665; Error (%) = (Reading − F) ÷ F × 100; Distance error = height read − gauge block size
Worked example
  1. 1

    2000 g: F = 2.000 kg × 9.80665 = 19.613 N

  2. 2
    Reading 19.580 N: error = (19.580 − 19.613) ÷ 19.613 × 100 = −0.17%
  3. 3

    Gauge block: 10.02 − 10.00 = +0.02 mm

ACCEPTANCE CRITERIA

Pass if

  • Force error typically within ± 0.5% of the applied load (or the manufacturer's specification) over the working range
  • Distance error within the manufacturer's specification, often a few hundredths of a millimetre
  • Speed within the specification stated by the supplier, if checked
IF IT FAILS

What to do

  • Stop using the load cell for reportable work.
  • Repeat the daily calibration and check the probe and platform are tight and clean.
  • If still out, check for overload damage and send the load cell for repair or recalibration.
  • Review results generated since the last passing check.
In LimsPulse · Calibration Management

LimsPulse schedules this calibration, records the result against limits, and blocks the instrument everywhere if it fails or goes overdue.

See how