Melt flow indexer

Melts a thermoplastic in a heated barrel and pushes it through a small standard die under a fixed weight. The mass (or volume) of polymer that flows out in 10 minutes is the melt flow rate, a quick measure of how easily the polymer flows.

  • Polymers & plastics
  • Packaging
  • Pipes & fittings
  • Automotive
  • Recycling
  • Chemicals
Photo of a Melt flow indexer
Photo: Miladfarhani, CC BY-SA 3.0 (cropped), via Wikimedia Commons
CALIBRATION

MFI tester calibration procedures

QUALIFICATION

Instrument qualification (DQ · IQ · OQ · PQ)

  1. 01

    DQ

    Define test conditions, temperature range, load set, Procedure A or B, and compliance with ISO 1133 and ASTM D1238.

  2. 02

    IQ

    Install on a level, vibration-free bench with ventilation; record serial numbers of barrel, die, piston and weights and their certificates.

  3. 03

    OQ

    Verify temperature at the die position, combined load masses, die dimensions, timer and encoder against traceable standards.

  4. 04

    PQ

    Test a reference polymer of known MFR and confirm the result falls within its stated range, then track it on a control chart.