Kjeldahl digestion & distillation unit
Digests a food sample in hot sulphuric acid to turn its organic nitrogen into ammonium, then distils the ammonia and titrates it. The nitrogen found, multiplied by a conversion factor, gives crude protein.
- Food & beverage
- Dairy
- Animal feed
- Agriculture
- Environmental
- Fertiliser

Kjeldahl calibration procedures
Nitrogen recovery check
Confirms that the distillation unit recovers all the ammonia (ammonium sulphate) and that digestion plus distillation recovers all the organic nitrogen (tryptophan or acetanilide).
Open procedureDigestion block temperature
Checks that the digestion block reaches its set temperature and that all tube positions are heated evenly.
Open procedureInstrument qualification (DQ · IQ · OQ · PQ)
- 01
DQ — Design qualification
Defines the needs: number of tubes, maximum block temperature, scrubber type, automatic or manual titration, software and data records.
- 02
IQ — Installation qualification
Confirms the units, utilities (power, cooling water, drain, exhaust), tubing and software are installed as specified, with serial numbers and manuals recorded.
- 03
OQ — Operational qualification
Checks block temperature accuracy and uniformity, alkali and water dosing, steam generation and titrator function against specifications.
- 04
PQ — Performance qualification
Shows the system gives correct results in routine use, for example nitrogen recovery with ammonium sulphate and an organic standard, plus a food reference material.
Keep every Kjeldahl in control: usage logs, calibration and maintenance schedules, and automatic blocking of overdue instruments.