Inductively Coupled Plasma Optical Emission Spectrometer (ICP-OES)
Excites atoms in an argon plasma at about 6000–10,000 K and measures the light each element emits. Measures many elements at once from trace to percent levels.
- Pharmaceutical
- Environmental
- Food & beverage
- Petrochemical
- Mining & metallurgy
- Chemicals

ICP-OES calibration procedures
Wavelength calibration
Aligns the spectrometer so that each element's emission line falls exactly where the software expects it, using a multi-element calibration solution.
Open procedureDetection limit check
Measures the noise of repeated blank readings and converts it into the lowest concentration the instrument can reliably detect for selected elements.
Open procedureInstrument qualification (DQ · IQ · OQ · PQ)
- 01
DQ — Design qualification
Defines elements, required detection limits, axial/radial view, sample types (aqueous or organic) and software compliance needs.
- 02
IQ — Installation qualification
Confirms installation, utilities (argon, exhaust, chiller, power), components, software version and documentation.
- 03
OQ — Operational qualification
Tests wavelength calibration, resolution, detection limits, short- and long-term stability and linearity.
- 04
PQ — Performance qualification
Shows ongoing fitness, e.g. daily performance checks and system suitability for the validated <233> procedure.
Keep every ICP-OES in control: usage logs, calibration and maintenance schedules, and automatic blocking of overdue instruments.