Inductively Coupled Plasma Optical Emission Spectrometer (ICP-OES)

Excites atoms in an argon plasma at about 6000–10,000 K and measures the light each element emits. Measures many elements at once from trace to percent levels.

  • Pharmaceutical
  • Environmental
  • Food & beverage
  • Petrochemical
  • Mining & metallurgy
  • Chemicals
Photo of a Inductively Coupled Plasma Optical Emission Spectrometer (ICP-OES)
Photo: Cyberwork 95, CC BY-SA 4.0 (cropped), via Wikimedia Commons
CALIBRATION

ICP-OES calibration procedures

QUALIFICATION

Instrument qualification (DQ · IQ · OQ · PQ)

  1. 01

    DQ — Design qualification

    Defines elements, required detection limits, axial/radial view, sample types (aqueous or organic) and software compliance needs.

  2. 02

    IQ — Installation qualification

    Confirms installation, utilities (argon, exhaust, chiller, power), components, software version and documentation.

  3. 03

    OQ — Operational qualification

    Tests wavelength calibration, resolution, detection limits, short- and long-term stability and linearity.

  4. 04

    PQ — Performance qualification

    Shows ongoing fitness, e.g. daily performance checks and system suitability for the validated <233> procedure.