Endotoxin (LAL) reader

An incubating photometer that holds LAL reactions at 37 °C and measures the change in absorbance or turbidity over time. Software turns the reaction onset time into an endotoxin concentration in EU/mL using a standard curve.

  • Pharmaceutical
  • Biotech
  • Medical devices
  • Dialysis and hospital pharmacy
  • Research laboratories
OVERVIEW

What is Endotoxin reader?

Bacterial endotoxins are lipopolysaccharides from the outer membrane of Gram-negative bacteria. They cause fever and shock if injected, so parenteral products and devices are tested for them. The test uses Limulus amebocyte lysate (LAL), a reagent from horseshoe crab blood that reacts with endotoxin in an enzyme cascade.

An endotoxin reader incubates LAL reactions at 37 ± 1 °C and reads them continuously. In the kinetic chromogenic method the cascade releases a yellow dye (p-nitroaniline) read at 405 nm. In the kinetic turbidimetric method the reaction forms a gel and the turbidity increases. The time taken to reach a set absorbance (the onset time) is shorter when there is more endotoxin.

Readers are either 96-well incubating microplate readers or multi-channel tube readers. Software builds a log-log standard curve of onset time against endotoxin concentration and calculates the endotoxin content of samples and the recovery of spiked samples. The gel-clot method does not need a reader, only a 37 °C heating block or water bath.

Used for

  • Bacterial endotoxins test (BET) of parenteral products, WFI and APIs
  • Testing of medical devices and container closures (rinse extracts)
  • Depyrogenation validation: endotoxin indicator recovery after dry heat
  • In-process monitoring of water systems and bioprocesses
  • Inhibition / enhancement testing during method validation
AT A GLANCE

How it works

  1. 01

    Prepare

    Standards, samples, spiked samples and negative controls are made in LAL reagent water using pyrogen-free tips and tubes.

  2. 02

    Pre-warm

    The plate or tubes are pre-incubated at 37 °C in the reader.

  3. 03

    Add LAL

    Reconstituted LAL reagent is added to every well, quickly and in a fixed order.

  4. 04

    Read kinetically

    Absorbance (405 nm) or turbidity is read at short intervals while held at 37 °C.

  5. 05

    Onset time

    Software records the time each well reaches the set optical density.

  6. 06

    Calculate

    Onset times are converted to EU/mL via the log-log standard curve; recovery of spikes is calculated.

Read the full working principle

Advantages

  • Quantitative result over a wide range (often 0.005–50 EU/mL)
  • Low limit of detection compared with gel-clot
  • Automated reading and calculation with audit trail
  • Many samples and replicates in one run (microplate)

Limitations

  • Results depend on accurate temperature control and pipetting
  • Products may inhibit or enhance the reaction and need dilution and validation
  • Glucans can give false positives with some LAL reagents
  • Software and reader need computer system validation
  • Reagent lots and endotoxin standards must be matched and qualified
CALIBRATION

1 calibration procedure

TEST METHODS

Methods that use Endotoxin reader

In LimsPulse · Calibration Management

Keep every Endotoxin reader in control: usage logs, calibration and maintenance schedules, and automatic blocking of overdue instruments.