Endotoxin (LAL) reader

An incubating photometer that holds LAL reactions at 37 °C and measures the change in absorbance or turbidity over time. Software turns the reaction onset time into an endotoxin concentration in EU/mL using a standard curve.

  • Pharmaceutical
  • Biotech
  • Medical devices
  • Dialysis and hospital pharmacy
  • Research laboratories
TROUBLESHOOTING

Common problems, causes and fixes

Standard curve |r| below 0.980
Likely causes
  • • Poor CSE dilution or vortexing
  • • Pipetting errors
  • • Temperature not uniform
What to do
  • • Prepare fresh standards with full vortexing
  • • Check pipette calibration and technique
  • • Check reader temperature calibration
Negative control shows endotoxin
Likely causes
  • • Contaminated LRW, tips or tubes
  • • Environmental contamination during set-up
What to do
  • • Use a new lot of LRW and consumables
  • • Clean the work area and repeat
PPC recovery below 50%
Likely causes
  • • Product inhibition (pH, chelators, protein binding)
  • • Spike added incorrectly
What to do
  • • Dilute further (not beyond the MVD)
  • • Adjust pH or use a dispersing buffer
  • • Repeat with correct spiking
PPC recovery above 200%
Likely causes
  • • Product enhancement
  • • Glucan interference
  • • Endotoxin contamination of the spiked sample
What to do
  • • Dilute further
  • • Use a glucan-blocking buffer or glucan-insensitive reagent
  • • Repeat with fresh consumables
Very short onset in all wells
Likely causes
  • • Contaminated LAL reagent
  • • Reader temperature too high
What to do
  • • Use a new reagent vial
  • • Check temperature