Endotoxin (LAL) reader

An incubating photometer that holds LAL reactions at 37 °C and measures the change in absorbance or turbidity over time. Software turns the reaction onset time into an endotoxin concentration in EU/mL using a standard curve.

  • Pharmaceutical
  • Biotech
  • Medical devices
  • Dialysis and hospital pharmacy
  • Research laboratories
LEARN & PRACTICE

Test your Endotoxin reader knowledge

Multiple-choice questions

Pick an answer to see whether you are right and why.

  1. Q1. Where do bacterial endotoxins come from?

    basic
  2. Q2. At what temperature are LAL reactions incubated?

    basic
  3. Q3. In the kinetic chromogenic method, what happens to the onset time as endotoxin concentration increases?

    intermediate
  4. Q4. At what wavelength is p-nitroaniline read in the chromogenic method?

    intermediate
  5. Q5. What is the USP <85> criterion for the standard curve correlation in kinetic methods?

    intermediate
  6. Q6. Positive product control recovery must be within:

    intermediate
  7. Q7. A product with a recommended dose of 10 mg/kg/h and K = 5 EU/kg has an endotoxin limit of:

    advanced

Viva questions

Think of your answer first, then open it to compare.

  1. 1. Name the three LAL techniques in USP <85> and how each detects endotoxin.

    Gel-clot: a firm gel forms at or above the lysate sensitivity (read by inverting the tube). Kinetic turbidimetric: turbidity increases as the clot forms, read over time. Kinetic chromogenic: a yellow product (p-nitroaniline) is released and read at 405 nm over time.

  2. 2. What is the maximum valid dilution and why does it matter?

    MVD = (endotoxin limit × concentration of the test solution) ÷ λ. It is the greatest dilution at which the test can still detect endotoxin at the limit. Samples are diluted to overcome interference, but never beyond the MVD.

  3. 3. The PPC recovery is 30%. What does it mean and what do you do?

    The product inhibits the reaction, so the sample result cannot be trusted. Repeat at a higher dilution (still within the MVD), or address the cause (pH adjustment, buffer, heat treatment if validated). Do not report the sample result from the invalid test.