Endotoxin (LAL) reader

An incubating photometer that holds LAL reactions at 37 °C and measures the change in absorbance or turbidity over time. Software turns the reaction onset time into an endotoxin concentration in EU/mL using a standard curve.

  • Pharmaceutical
  • Biotech
  • Medical devices
  • Dialysis and hospital pharmacy
  • Research laboratories
WORKING PRINCIPLE

How Endotoxin reader works

Endotoxin activates Factor C in the LAL, which starts a protease cascade. In chromogenic reagents the final enzyme cuts a synthetic substrate and releases yellow p-nitroaniline (read at 405 nm). In turbidimetric reagents the enzyme converts coagulogen to coagulin, which makes the solution turbid. The reader measures each well or tube over time; the onset time falls as endotoxin increases, and log onset time vs log concentration is linear.

FLOW

The path of a sample

  1. 01

    Prepare

    Standards, samples, spiked samples and negative controls are made in LAL reagent water using pyrogen-free tips and tubes.

  2. 02

    Pre-warm

    The plate or tubes are pre-incubated at 37 °C in the reader.

  3. 03

    Add LAL

    Reconstituted LAL reagent is added to every well, quickly and in a fixed order.

  4. 04

    Read kinetically

    Absorbance (405 nm) or turbidity is read at short intervals while held at 37 °C.

  5. 05

    Onset time

    Software records the time each well reaches the set optical density.

  6. 06

    Calculate

    Onset times are converted to EU/mL via the log-log standard curve; recovery of spikes is calculated.

COMPONENTS

Main parts and what they do

1

Incubation block

Holds the microplate or tubes at 37 ± 1 °C.

2

Light source and filter (405 nm)

Provides light at the wavelength of p-nitroaniline for chromogenic tests.

3

Photodetectors

Measure transmitted light through each well or tube.

4

Plate carrier / tube wells

Position each reaction in the light path.

5

Temperature sensors

Monitor and control the block temperature.

6

Endotoxin software

Controls reading, builds the standard curve, calculates results and keeps an audit trail.

KEY CONCEPTS

Terms every analyst should know

Endotoxin unit (EU)

Unit of endotoxin activity defined against the reference standard endotoxin (RSE). Control standard endotoxin (CSE) is calibrated against it.

Onset time

Time for the reaction to reach a set absorbance change. More endotoxin gives a faster reaction and a shorter onset time.

Lysate sensitivity (λ)

For kinetic methods, the lowest concentration on the standard curve; for gel-clot, the labelled sensitivity of the lysate (e.g. 0.03 EU/mL).

Positive product control (PPC)

Sample spiked with a known amount of endotoxin. Recovery shows that the sample does not inhibit or enhance the reaction.

Inhibition and enhancement

Product factors such as pH, chelators, proteins or high salt can slow or speed the reaction. Dilution up to the MVD is the usual remedy.

Temperature sensitivity

The enzyme cascade rate depends on temperature, so the reader must hold 37 ± 1 °C evenly across all wells.

In LimsPulse · Calibration Management

Keep every Endotoxin reader in control: usage logs, calibration and maintenance schedules, and automatic blocking of overdue instruments.