Endotoxin (LAL) reader
An incubating photometer that holds LAL reactions at 37 °C and measures the change in absorbance or turbidity over time. Software turns the reaction onset time into an endotoxin concentration in EU/mL using a standard curve.
- Pharmaceutical
- Biotech
- Medical devices
- Dialysis and hospital pharmacy
- Research laboratories
Endotoxin reader calibration procedures
Instrument qualification (DQ · IQ · OQ · PQ)
- 01
DQ — Design qualification
Defines the reader type (plate or tube), methods (chromogenic, turbidimetric), wavelength, capacity, software and data integrity needs.
- 02
IQ — Installation qualification
Records the reader, computer, software version, user roles and calibration certificate.
- 03
OQ — Operational qualification
Verifies temperature accuracy and uniformity, photometric accuracy and linearity, read timing and software calculations.
- 04
PQ — Performance qualification
Runs standard curves and spiked samples to show the system meets USP <85> criteria, and qualifies each analyst with the method.
Keep every Endotoxin reader in control: usage logs, calibration and maintenance schedules, and automatic blocking of overdue instruments.