Endotoxin reader · Calibration procedure

Temperature and photometric checks

Verifies that the reader holds all wells or tubes at 37 ± 1 °C and that its absorbance readings at 405 nm (or turbidity readings) agree with certified reference filters or plates.

PURPOSE

Why this is checked

The LAL reaction rate depends strongly on temperature, and the onset time depends on the measured optical density. Uneven heating or a photometric error shifts onset times and so changes the endotoxin result. These checks confirm the reader is fit for USP <85> kinetic methods.

Frequency

Typically every 6–12 months, and after repair or relocation, as defined in your SOP.

Learn how the Endotoxin reader works
YOU WILL NEED

Standards & tools

  • Calibrated temperature probe or temperature verification plate traceable to national standards
  • Certified neutral-density filter plate or absorbance reference standards at 405 nm
  • Manufacturer's verification software or test plate, where provided
BEFORE YOU START

Prerequisites

  • Reader switched on and at temperature for the manufacturer's warm-up time
  • Optical path clean
PROCEDURE

Step by step

  1. 1Temperature: place the calibrated probe(s) in representative wells (corners and centre) or tube positions with LRW and let them equilibrate.
  2. 2Record the temperature at each position and compare with 37 °C.
  3. 3Photometric: read the certified filter plate at 405 nm and record each value.
  4. 4Calculate the error against the certified value for each filter.
  5. 5For tube readers, use the manufacturer's optical check and record the result.
RECORD

Example observation table

CheckReferenceMeasuredError
Temperature, well A137.0 °C36.8 °C−0.2 °C
Temperature, well A1237.0 °C37.1 °C+0.1 °C
Temperature, well D6 (centre)37.0 °C37.3 °C+0.3 °C
Temperature, well H137.0 °C36.9 °C−0.1 °C
Temperature, well H1237.0 °C37.0 °C0.0 °C
OD at 405 nm, filter 10.5000.503+0.6%
OD at 405 nm, filter 21.0001.006+0.6%
OD at 405 nm, filter 32.0002.012+0.6%
CALCULATION

How to calculate the result

Temperature error = Measured − 37.0 °C; Photometric error (%) = (Measured OD − Certified OD) ÷ Certified OD × 100
  1. Highest temperature 37.3 °C and lowest 36.8 °C: all within 36–38 °C
  2. Filter 2: (1.006 − 1.000) ÷ 1.000 × 100 = +0.6%
  3. Filter 3: (2.012 − 2.000) ÷ 2.000 × 100 = +0.6%
  4. Within a typical tolerance of ±1% (or the filter certificate tolerance): Pass
ACCEPTANCE CRITERIA

Pass if

  • All positions at 37 ± 1 °C (the incubation temperature in USP <85>).
  • Absorbance within the tolerance given by the manufacturer or the filter certificate, often about ±1% plus a small fixed amount (e.g. 0.010 OD), as defined in your SOP.
IF IT FAILS

What to do

  • Label the reader 'Not for use'.
  • Have the manufacturer adjust or repair the heater or optics, then repeat the checks.
  • Assess endotoxin results since the last passing check with QA.