CALIBRATION

Conductivity meter calibration procedures

QUALIFICATION

Instrument qualification (DQ · IQ · OQ · PQ)

  1. 01

    DQ — Design qualification

    Defines range and resolution (0.1 µS/cm or better for pharma water), cell constant, temperature measurement, ability to show non-compensated values and data integrity needs.

  2. 02

    IQ — Installation qualification

    Records meter, cell and sensor details, firmware and installation (offline bench or online in the water loop).

  3. 03

    OQ — Operational qualification

    Checks cell constant with a certified standard, meter accuracy with precision resistors where required, and temperature sensor accuracy.

  4. 04

    PQ — Performance qualification

    Shows consistent performance in routine use, for example periodic standard checks and comparison of online and offline readings.