LimsPulse
Conductivity meter · Calibration procedure

Conductivity standard check

Measures one or more certified conductivity standards as samples to verify the meter and cell read correctly after calibration.

Photo of a Conductivity meter
Photo: Nuno Nogueira (Nmnogueira), CC BY-SA 2.5 (cropped), via Wikimedia Commons
PURPOSE

Why this is checked

A verification with an independent standard confirms the whole system, including cell constant, temperature handling and electronics, gives correct conductivity values.

Frequency

Before use or weekly / monthly, and after each cell-constant calibration, as defined in your SOP.

Learn how the Conductivity meter works
YOU WILL NEED

Standards & tools

  • Certified conductivity standards from a different lot than the one used for calibration, e.g. 146.9 µS/cm and 1413 µS/cm at 25 °C
  • Calibrated reference thermometer (for the temperature check)
BEFORE YOU START

Prerequisites

  • Cell constant up to date in the meter
  • Temperature compensation set to the reference temperature of the standard (25 °C) with the coefficient stated on the certificate, or measure at exactly 25 °C
PROCEDURE

Step by step

  1. 1Rinse the cell with portions of the first standard and discard them.
  2. 2Measure a fresh portion and record conductivity and temperature when stable.
  3. 3Repeat with the second standard.
  4. 4Compare the temperature reading with the reference thermometer in the same beaker.
  5. 5Calculate the % error for each standard.
RECORD

Example observation table

Standard (µS/cm at 25 °C)Observed (µS/cm)% errorMeter temp. (°C)Reference temp. (°C)
146.9147.8+0.61%25.125.0
14131408−0.35%25.025.0
CALCULATION

How to calculate the result

% error = (Observed − Certified value) ÷ Certified value × 100
  1. 146.9 µS/cm standard reads 147.8 µS/cm
  2. % error = (147.8 − 146.9) ÷ 146.9 × 100 = +0.61%
ACCEPTANCE CRITERIA

Pass if

  • Observed value within the tolerance in your SOP, commonly within ±2% (some labs use ±3% or the certificate's uncertainty plus meter accuracy).
  • Temperature within ±0.5 °C of the reference (USP <645> requires temperature measurement good to at least ±2 °C).
IF IT FAILS

What to do

  • Repeat with a fresh portion of standard and a well-rinsed cell.
  • Re-determine the cell constant and repeat the check.
  • If it still fails, take the meter out of use and investigate results since the last pass.
MORE

Other Conductivity meter calibrations

In LimsPulse · Calibration Management

LimsPulse schedules this calibration, records the result against limits, and blocks the instrument everywhere if it fails or goes overdue.