Cloud and Pour Point Apparatus
Cools a fuel or oil in a standard glass test jar through a series of cold baths and examines it at set intervals. The cloud point is where wax crystals first appear; the pour point is the lowest temperature at which the sample still flows.
- Petroleum refining
- Lubricant blending
- Fuel distribution in cold climates
- Crude oil production and pipelines
- Independent testing laboratories
Cloud & pour point tester calibration procedures
Instrument qualification (DQ · IQ · OQ · PQ)
- 01
DQ — Design qualification
Defines methods (ASTM D2500, D97, ISO 3015/3016, IS 1448 Part 10), lowest temperature needed, number of baths, manual or automatic detection, and coolant type.
- 02
IQ — Installation qualification
Confirms installation, ventilation of refrigeration units, coolant filled, thermometers and jars supplied with certificates, documents recorded.
- 03
OQ — Operational qualification
Verifies each bath reaches and holds its set temperature, thermometer accuracy, and for automatic units the cooling rate and detection function.
- 04
PQ — Performance qualification
Tests reference or check samples with known cloud and pour points and shows agreement within the method precision; automatic units are compared with the manual method.
Keep every Cloud & pour point tester in control: usage logs, calibration and maintenance schedules, and automatic blocking of overdue instruments.