Cloud & pour point tester · Calibration procedure

Thermometer verification

Checks the cloud and pour point thermometers (or digital probes) at the ice point and against a calibrated reference thermometer at low temperatures.

Illustration of Cloud & pour point tester
PURPOSE

Why this is checked

Cloud and pour points are read directly from the test thermometer, so any thermometer error goes straight into the result. Liquid-in-glass thermometers can change through glass ageing or a separated column; digital probes can drift. A regular check at the ice point and at low temperatures confirms they read correctly.

Frequency

Typically an ice-point check every 6 months and a multi-point comparison annually, or per SOP; also after any suspected damage.

Learn how the Cloud & pour point tester works
YOU WILL NEED

Standards & tools

  • Ice point bath made from crushed ice of distilled water and distilled water
  • Calibrated reference thermometer traceable to national standards
  • Stirred low-temperature bath able to reach the lowest working temperature
BEFORE YOU START

Prerequisites

  • Visual check: no separated column, cracks or illegible markings
  • Reference thermometer in date
PROCEDURE

Step by step

  1. 1Prepare an ice point bath and immerse the thermometer to its specified immersion depth; read after it stabilises.
  2. 2Place the test thermometer next to the reference in the stirred low-temperature bath.
  3. 3Set the bath to each check temperature (for example −20 °C and −40 °C) and record both readings once stable.
  4. 4Calculate the error at each point and update the correction table if your SOP uses corrections.
RECORD

Example observation table

PointReference (°C)Test thermometer (°C)Error (°C)
Ice point0.000.2+0.2
Low 1−20.00−19.6+0.4
Low 2−40.00−39.5+0.5
CALCULATION

How to calculate the result

Formula
Error = test reading − reference reading; Correction = − error
Worked example
  1. 1

    Ice point: 0.2 − 0.00 = +0.2 °C.

  2. 2

    −20 °C: −19.6 − (−20.00) = +0.4 °C.

  3. 3

    −40 °C: −39.5 − (−40.00) = +0.5 °C.

  4. 4

    With a limit of ±0.5 °C (half of a 1 °C graduation, illustrative), all points pass; the −40 °C point is at the limit and should be watched.

ACCEPTANCE CRITERIA

Pass if

  • Error within the limit in your SOP (commonly about half a scale division, for example ±0.5 °C for a thermometer graduated in 1 °C).
  • No separated column or damage.
IF IT FAILS

What to do

  • Try to rejoin a separated column as the manufacturer advises, then re-check.
  • Apply a documented correction if the SOP allows; otherwise replace the thermometer.
  • Assess cloud and pour point results since the last good check.