Capillary electrophoresis system
Separates charged molecules in a thin fused-silica capillary under a high voltage (typically up to 30 kV). Used in biologics QC for size purity (CE-SDS), charge variants (cIEF, CZE) and glycan analysis.
- Biotechnology
- Biopharmaceutical
- Pharmaceutical
- Genomics
- Forensic science
- Academic research

CE calibration procedures
Instrument qualification (DQ · IQ · OQ · PQ)
- 01
DQ
Define separation modes, detectors, voltage range, temperature control range, autosampler cooling and software compliance needs.
- 02
IQ
Install, check electrical safety and cooling, install software and record serial numbers and capillary details.
- 03
OQ
Verify capillary temperature, voltage and current accuracy, injection precision, detector wavelength accuracy, noise and drift.
- 04
PQ
Run a reference standard by the routine method (for example CE-SDS of a reference antibody) and confirm migration time, resolution and area precision meet set criteria.
Keep every CE in control: usage logs, calibration and maintenance schedules, and automatic blocking of overdue instruments.