CE · Calibration procedure

Temperature and voltage verification

Checks the capillary cartridge temperature and the high-voltage output of the CE system against calibrated tools or the manufacturer's verification kit.

Photo of a CE
Photo: Mark Pellegrini, CC BY-SA 3.0 (cropped), via Wikimedia Commons
PURPOSE

Why this is checked

Mobility rises by about 2% for every 1 °C, so temperature errors shift migration times and can change resolution. Voltage errors change field strength and therefore migration time and heating. Both must be within specification for validated CE methods to transfer between instruments.

Frequency

Typically every 6–12 months and after repair, as defined in your SOP.

Learn how the CE works
YOU WILL NEED

Standards & tools

  • Calibrated temperature probe or manufacturer's cartridge temperature verification tool
  • Calibrated high-voltage measurement device or manufacturer's voltage verification kit (often a high-voltage divider)
  • Test capillary and buffer as described in the kit, if a current check is included
BEFORE YOU START

Prerequisites

  • Only trained engineers or analysts perform the high-voltage check
  • Interlocks working; covers closed during measurement
  • Tools within calibration date
PROCEDURE

Step by step

  1. 1Fit the temperature verification tool in the cartridge position.
  2. 2Set the cartridge to several temperatures (for example 20, 25 and 40 °C), allow it to stabilise and record measured values.
  3. 3Install the voltage verification kit as instructed.
  4. 4Apply several voltages (for example 10, 20 and 30 kV) and record measured values.
  5. 5Where required, run buffer at increasing voltages and plot current against voltage (Ohm's law plot) to confirm heat removal.
  6. 6Calculate differences from set values.
RECORD

Example observation table

ParameterSetMeasuredDifferenceResult
Temperature20 °C20.3 °C+0.3 °CPass
Temperature25 °C25.2 °C+0.2 °CPass
Temperature40 °C39.6 °C−0.4 °CPass
Voltage10 kV10.05 kV+0.5%Pass
Voltage20 kV19.88 kV−0.6%Pass
Voltage30 kV29.85 kV−0.5%Pass
CALCULATION

How to calculate the result

Formula
Temperature difference = measured − set; Voltage error (%) = (measured − set) ÷ set × 100
Worked example
  1. 1

    40 °C: 39.6 − 40.0 = −0.4 °C

  2. 2
    20 kV: (19.88 − 20) ÷ 20 × 100 = −0.6%
  3. 3
    30 kV: (29.85 − 30) ÷ 30 × 100 = −0.5%
ACCEPTANCE CRITERIA

Pass if

  • Cartridge temperature typically within about ±1 °C of set point (or the manufacturer's specification)
  • Voltage typically within about ±1–2% of set value (or the manufacturer's specification)
  • Ohm's law plot linear over the method's voltage range, where tested
IF IT FAILS

What to do

  • Label the system out of use for GMP analyses.
  • Check coolant level and cartridge seating for temperature failures; call service for adjustment.
  • Call service for high-voltage supply repair or calibration.
  • Assess analyses run since the last passing check (migration time and resolution trends).