Thin-Layer Chromatography (TLC)

Separates compounds as a solvent climbs up a thin layer of adsorbent on a glass, aluminium or plastic plate. Simple and cheap; used in pharma QC mainly for identification and limit tests for related substances.

  • Pharmaceutical
  • Herbal & natural products
  • Chemicals
  • Forensic
  • Education
Photo of a Thin-Layer Chromatography (TLC)
Photo: Daniel Dróżdż, CC BY 4.0 (cropped), via Wikimedia Commons
TROUBLESHOOTING

Common problems, causes and fixes

Streaking or tailing spots
Likely causes
  • • Sample overloaded
  • • Sample too polar or ionised
  • • Spot not dried before development
What to do
  • • Apply less sample or dilute it
  • • Add a little acid or base (e.g. acetic acid or ammonia) to the mobile phase if the method allows
  • • Dry spots fully before developing
Uneven or curved solvent front
Likely causes
  • • Chamber not saturated
  • • Plate touching the filter paper or wall
  • • Uneven layer or damaged plate edge
What to do
  • • Saturate for longer with a paper lining
  • • Position the plate carefully
  • • Use a new plate
All spots near the origin
Likely causes
  • • Mobile phase not polar enough
  • • Wrong plate type
What to do
  • • Check mobile-phase composition against the method
  • • Confirm plate type and activation
All spots near the solvent front
Likely causes
  • • Mobile phase too polar
  • • Plate deactivated by humidity
What to do
  • • Check mobile-phase composition
  • • Activate plates and store them dry
Rf values differ from expected
Likely causes
  • • Temperature or humidity differences
  • • Different plate batch
  • • Old or evaporated mobile phase
What to do
  • • Always compare with a reference on the same plate
  • • Use fresh mobile phase each time
  • • Control saturation time
Spots not visible
Likely causes
  • • Compound does not absorb at 254 nm
  • • Too little sample applied
  • • Wrong visualisation reagent
What to do
  • • Use a suitable chemical reagent or iodine
  • • Increase application volume within the method
  • • Check the reagent preparation and heating step