LimsPulse
Titration & electrochemistry

Potentiometric Titrator (Autotitrator)

Adds a titrant from a motorised burette while an electrode measures the potential of the solution, and finds the equivalence point from the change in potential. Used for assays, standardisation and many limit tests.

  • Pharmaceutical
  • Chemicals
  • Food & beverage
  • Environmental
  • Metal finishing
Photo of a Potentiometric Titrator (Autotitrator)
Photo: User:FA2010, Public domain (cropped), via Wikimedia Commons
TROUBLESHOOTING

Common problems, causes and fixes

No clear equivalence point
Likely causes
  • • Wrong electrode for the titration type
  • • Very weak acid or base in the chosen solvent
  • • Sample mass too small
What to do
  • • Use the electrode the method specifies
  • • Use a non-aqueous solvent system if the method allows
  • • Increase sample mass to consume more titrant
Noisy or drifting potential
Likely causes
  • • Blocked or dirty diaphragm
  • • Low reference electrolyte
  • • Static or poor earthing in non-aqueous solvents
  • • Air bubble under the glass membrane
What to do
  • • Clean the diaphragm or replace the electrode
  • • Refill electrolyte and open the refill port during titration
  • • Use an electrode suited to non-aqueous media
  • • Shake the electrode to remove the bubble
Results high or low compared with expected
Likely causes
  • • Old or wrong titre of the titrant
  • • Air bubbles in the burette
  • • Blank not subtracted
  • • Wrong sample mass entry
What to do
  • • Restandardise the titrant
  • • Flush the burette
  • • Run and apply the blank
  • • Check the mass transfer from the balance
Slow electrode response
Likely causes
  • • Ageing glass membrane
  • • Protein or fat film on the membrane
  • • Electrode stored dry
What to do
  • • Clean or recondition the membrane as per supplier
  • • Replace the electrode if slope and response remain poor
Two or more equivalence points detected
Likely causes
  • • Polyprotic analyte or impurity
  • • Equivalence point recognition set too sensitive
What to do
  • • Confirm which point the method uses
  • • Adjust the recognition threshold or window within the validated range
Leaking burette tip or dripping
Likely causes
  • • Worn piston seal or valve
  • • Loose tubing connection
What to do
  • • Tighten fittings
  • • Replace the seal or service the burette unit
In LimsPulse · Calibration Management

Keep every Potentiometric titrator in control: usage logs, calibration and maintenance schedules, and automatic blocking of overdue instruments.

See how