Particulate Matter (PM10 / PM2.5) Sampler
Draws a known volume of ambient air through a size-selective inlet and a filter so that the mass of PM10 or PM2.5 particles collected can be weighed and expressed as µg/m³.
- Environmental monitoring agencies and pollution control boards
- Environmental consultancies
- Mining, cement, power and steel
- Research institutions
- Construction and infrastructure projects

PM sampler calibration procedures
Instrument qualification (DQ · IQ · OQ · PQ)
- 01
Design qualification (DQ)
Select a sampler design (high or low volume, PM10 or PM2.5) that meets the requirements of the applicable national standard or reference method.
- 02
Installation qualification (IQ)
Record serial numbers, install at a site that meets siting guidance, verify power, housing and data logging.
- 03
Operational qualification (OQ)
Leak check, multi-point flow calibration against a certified transfer standard, and checks of temperature, pressure and timer.
- 04
Performance qualification (PQ)
Run collocated samplers or field blanks and show precision and blank levels meet the lab's quality objectives.
Keep every PM sampler in control: usage logs, calibration and maintenance schedules, and automatic blocking of overdue instruments.