LimsPulse
PM sampler · Calibration procedure

Flow rate calibration

Checks and adjusts the sampler's flow rate against a calibrated flow transfer standard so the inlet cut-point and the sampled air volume are correct.

Photo of a PM sampler
Photo: Srini297, CC BY-SA 4.0 (cropped), via Wikimedia Commons
PURPOSE

Why this is checked

The PM concentration equals filter mass gain divided by air volume, so a flow error causes an equal error in the result. Flow also controls the inlet cut-point: if the flow is wrong, the sampler collects the wrong particle size range.

Frequency

Typically a single-point flow check monthly (or as required by the monitoring programme) and a multi-point calibration on installation, after repairs and at least annually, as defined in your SOP.

Learn how the PM sampler works
YOU WILL NEED

Standards & tools

  • Certified flow transfer standard (for example a calibrated orifice for high-volume samplers, or a volumetric or mass flow calibrator for low-volume samplers), traceable to national standards
  • Calibrated thermometer and barometer for ambient temperature and pressure
BEFORE YOU START

Prerequisites

  • Sampler leak check passed
  • A clean filter (or the type of filter used in sampling) installed
  • Ambient temperature and pressure sensors of the sampler checked
PROCEDURE

Step by step

  1. 1Install a clean filter and connect the transfer standard to the inlet using the adapter.
  2. 2Run the sampler at its design flow (for example 16.67 L/min for a low-volume PM2.5 sampler) and allow it to stabilise.
  3. 3Record the sampler's indicated flow and the transfer standard flow, with ambient temperature and pressure.
  4. 4For a multi-point calibration, repeat at several flow points around the design flow and fit the calibration line.
  5. 5Calculate the difference between indicated and standard flow, and between standard flow and design flow.
  6. 6Adjust the sampler calibration if needed and repeat the check.
RECORD

Example observation table

CheckIndicated flow (L/min)Transfer standard flow (L/min)Indicated vs standard (%)Standard vs design 16.67 L/min (%)
As found16.6716.41+1.58−1.56
As left (after adjustment)16.6716.52+0.91−0.90
CALCULATION

How to calculate the result

Formula
Flow difference (%) = (indicated − standard) ÷ standard × 100; Design flow error (%) = (standard − design) ÷ design × 100
Worked example
  1. 1
    As found: (16.67 − 16.41) ÷ 16.41 × 100 = +1.58%
  2. 2
    Design error: (16.41 − 16.67) ÷ 16.67 × 100 = −1.56%
  3. 3
    Both within typical limits, but adjusted to bring closer: (16.67 − 16.52) ÷ 16.52 × 100 = +0.91%
ACCEPTANCE CRITERIA

Pass if

  • Indicated flow typically within ±4% of the transfer standard flow (as given in common reference method guidance; use your SOP)
  • Measured flow typically within ±5% of the design flow rate
  • Leak check passes the manufacturer's criterion
IF IT FAILS

What to do

  • Do not start a sampling run until the flow passes.
  • Check for leaks, a blocked inlet or a wrong filter, then repeat.
  • Recalibrate the flow sensor or service the pump.
  • Assess or invalidate runs since the last passing check where the flow error affects data.
In LimsPulse · Calibration Management

LimsPulse schedules this calibration, records the result against limits, and blocks the instrument everywhere if it fails or goes overdue.

See how