Microplate reader

Measures absorbance, fluorescence or luminescence in the wells of 96- or 384-well plates. It is the read-out instrument for ELISA, protein assays, cell-based assays and enzyme kinetics.

  • Biotechnology
  • Biopharmaceutical
  • Clinical research
  • Academic research
  • Food (allergen ELISA)
  • Environmental (toxicity assays)
Photo of a Microplate reader
Photo: Pdcook, CC BY-SA 3.0 (cropped), via Wikimedia Commons
TROUBLESHOOTING

Common problems, causes and fixes

Readings unstable or drift
Likely causes
  • • Lamp not warmed up or ageing
  • • Condensation on plate (cold plate in warm reader)
What to do
  • • Allow warm-up; replace lamp if energy low
  • • Equilibrate plates to room temperature before reading
Random outlier wells
Likely causes
  • • Bubbles or particles
  • • Scratched or dirty plate bottom
  • • Pipetting error
What to do
  • • Remove bubbles and centrifuge
  • • Use new plates and handle by the edges
  • • Check pipettes and technique
Outer wells read higher or lower
Likely causes
  • • Evaporation during incubation
  • • Temperature gradient across the plate
What to do
  • • Use plate seals and humidified incubation
  • • Pre-warm reagents and plates; avoid stacking plates
Calibration plate values out of tolerance
Likely causes
  • • Dirty or damaged filters
  • • Lamp or detector fault
  • • Expired calibration plate certificate
What to do
  • • Clean filters as the supplier advises
  • • Service the reader
  • • Recertify or replace the calibration plate
Standard curve flattens at high concentration
Likely causes
  • • Above the photometric linear range
  • • Substrate exhausted in the assay
What to do
  • • Dilute samples into range
  • • Shorten development time or adjust the assay
High background in fluorescence
Likely causes
  • • Wrong plate colour (clear plate)
  • • Autofluorescence of medium or plate
  • • Gain set too high
What to do
  • • Use black plates
  • • Use phenol-red-free medium or blank-correct
  • • Optimise gain on the highest standard