LimsPulse
Bioanalytical

Microplate reader

Measures absorbance, fluorescence or luminescence in the wells of 96- or 384-well plates. It is the read-out instrument for ELISA, protein assays, cell-based assays and enzyme kinetics.

  • Biotechnology
  • Biopharmaceutical
  • Clinical research
  • Academic research
  • Food (allergen ELISA)
  • Environmental (toxicity assays)
Photo of a Microplate reader
Photo: Pdcook, CC BY-SA 3.0 (cropped), via Wikimedia Commons
Follow your site SOP
This is a typical operating sequence for learning. Your instrument model and approved SOP take precedence.
STAGE 1

Before you start

  1. 1Check the reader is within its calibration (absorbance, linearity, wavelength) due date.
  2. 2Switch on and allow the lamp and incubator (if used) to stabilise as the manual says.
  3. 3Select or open the validated protocol; check wavelengths, read mode, shaking and plate type.
  • Use the plate type the protocol was validated with (clear flat-bottom for absorbance, black for fluorescence, white for luminescence).
STAGE 2

Reading

  1. 1Wipe the plate bottom and check for bubbles; pop bubbles with a clean needle or brief centrifugation.
  2. 2Remove the plate lid or seal unless the protocol says otherwise.
  3. 3Load the plate with well A1 in the correct position.
  4. 4Start the read and check the raw data for out-of-range or flagged wells.
  • Do not read plates with liquid on the outside; spills can enter the optics.
STAGE 3

Data review

  1. 1Confirm blanks, standards and controls meet the method criteria.
  2. 2Check the curve fit (r² or goodness-of-fit, back-calculated standards).
  3. 3Export or save results with the raw data and audit trail.
STAGE 4

Shutdown

  1. 1Remove the plate and close the carrier.
  2. 2Wipe any spills from the carrier with a lint-free cloth.
  3. 3Complete the logbook.
In LimsPulse · Calibration Management

Keep every Plate reader in control: usage logs, calibration and maintenance schedules, and automatic blocking of overdue instruments.

See how