Microbial air sampler (active air sampler)
Draws a known volume of air across an agar plate so that airborne microorganisms are collected and can be counted as cfu per m³. Used for active viable air monitoring of clean rooms, isolators and LAF units.
- Pharmaceutical
- Biotech
- Hospitals and compounding pharmacies
- Medical devices
- Food & beverage

Air sampler calibration procedures
Instrument qualification (DQ · IQ · OQ · PQ)
- 01
DQ — Design qualification
Defines flow rate, sampling volumes, head type and d50, suitability for isolators, battery life, data storage and cleaning compatibility.
- 02
IQ — Installation qualification
Records the unit, heads, firmware, calibration certificate and manuals.
- 03
OQ — Operational qualification
Verifies flow rate and volume accuracy with a calibrated flow meter, timer accuracy, alarms and data recording.
- 04
PQ — Performance qualification
Shows the sampler, heads and media recover organisms in routine use, and compares results with an established sampler where one is being replaced.
Keep every Air sampler in control: usage logs, calibration and maintenance schedules, and automatic blocking of overdue instruments.