Microbial air sampler (active air sampler)

Draws a known volume of air across an agar plate so that airborne microorganisms are collected and can be counted as cfu per m³. Used for active viable air monitoring of clean rooms, isolators and LAF units.

  • Pharmaceutical
  • Biotech
  • Hospitals and compounding pharmacies
  • Medical devices
  • Food & beverage
Photo of a Microbial air sampler (active air sampler)
Photo: U.S. Marine Corps photo by Cpl. Earik Barton, Public domain (cropped), via Wikimedia Commons
TROUBLESHOOTING

Common problems, causes and fixes

Flow error alarm during sampling
Likely causes
  • • Blocked head holes
  • • Fan or sensor fault
  • • Low battery
What to do
  • • Clean the head and check holes
  • • Charge or replace the battery
  • • Send for service and recalibration
Unexpected growth on Grade A air samples
Likely causes
  • • Contaminated head or poor plate handling
  • • Real excursion in the area
  • • Sampler placed in turbulent air
What to do
  • • Review handling and head sanitisation
  • • Start an investigation with particle data and other EM results
  • • Identify the isolate and check placement
Agar surface dried or cracked after sampling
Likely causes
  • • Large sample volume in one plate
  • • Low agar fill volume
What to do
  • • Split the volume across plates or use thicker plates
  • • Validate the maximum volume per plate
Calibration shows flow outside tolerance
Likely causes
  • • Sensor drift
  • • Worn fan or leaking seal
What to do
  • • Adjust and recalibrate
  • • Assess impact on results since the last passing calibration