PREVENTIVE MAINTENANCE

Maintenance checklist

Daily / each use

  • Check for leaks at fittings
  • Note system pressure and compare with the normal range
  • Check seal-wash solvent level
  • Discard used mobile phase; never top up old mobile phase

Weekly

  • Clean solvent reservoirs
  • Check inlet filters for blockage
  • Check autosampler needle and seat for deposits

Monthly / quarterly

  • Replace inlet filters
  • Check lamp hours and energy
  • Clean or replace check valves if pressure ripple increases

Six-monthly / annual

  • Preventive maintenance by a trained engineer (pump seals, rotor seal, needle, lamp as needed)
  • Calibration of pump, injector, detector and column oven
SAFETY

Safety precautions

Before you work on the HPLC
  • Organic solvents such as methanol and acetonitrile are flammable and toxic: handle them in a fume hood and keep them away from ignition sources.
  • The system runs at high pressure: never loosen fittings while flow is on.
  • Do not look directly at the UV lamp.
  • Collect solvent waste in labelled, closed containers and dispose of it as per site procedure.
  • Wear safety glasses, gloves and a lab coat when preparing mobile phases.