Couples a gas chromatograph to a mass spectrometer so each volatile compound is separated and then identified from its fragmentation pattern. Used for unknown volatile impurities, nitrosamines, residual solvent confirmation and extractables.

  • Pharmaceutical
  • Environmental
  • Forensic and toxicology
  • Food & flavour
  • Petrochemical
  • Packaging
Photo of a Gas Chromatography–Mass Spectrometry (GC-MS)
Photo: Polimerek, CC BY-SA 4.0 (cropped), via Wikimedia Commons
CALIBRATION

GC-MS calibration procedures

QUALIFICATION

Instrument qualification (DQ · IQ · OQ · PQ)

  1. 01

    DQ — Design qualification

    Defines mass range, sensitivity, scan speed, ionisation modes, sample introduction (liquid, headspace) and software / library needs.

  2. 02

    IQ — Installation qualification

    Confirms installation, gas supplies, vacuum pump-down, exhaust, software and library versions.

  3. 03

    OQ — Operational qualification

    Tests GC functions (oven temperature, flow) and MS functions: mass calibration, resolution, sensitivity (S/N for a test compound such as octafluoronaphthalene) and scan performance.

  4. 04

    PQ — Performance qualification

    Routine tune checks and a system suitability standard show the system stays fit for its methods.