LimsPulse
Chromatography

Gas Chromatograph (GC)

Separates volatile compounds in a heated capillary column using a flowing inert gas and measures each one as it leaves, usually with a flame ionisation detector. Used in pharma QC for residual solvents, volatile impurities and some assays.

  • Pharmaceutical
  • Petrochemical
  • Food & flavour
  • Environmental
  • Forensic
  • Chemicals
Photo of a Gas Chromatograph (GC)
Photo: Sarka Na kopci, CC BY-SA 4.0 (cropped), via Wikimedia Commons
TROUBLESHOOTING

Common problems, causes and fixes

No peaks
Likely causes
  • • FID flame not lit or blown out
  • • Syringe blocked or empty vial
  • • Column broken or not connected
  • • Carrier gas cylinder empty
What to do
  • • Check hydrogen and air flows and re-ignite
  • • Replace the syringe and check vial volume
  • • Inspect the column and its connections
  • • Replace the cylinder and purge the lines
Ghost peaks or rising baseline
Likely causes
  • • Septum bleed (siloxane peaks)
  • • Carry-over from a previous injection
  • • Contaminated carrier gas or saturated traps
  • • Column bleed at high temperature
What to do
  • • Use a low-bleed septum and replace it
  • • Increase syringe washes and run blanks
  • • Replace gas traps or cylinder
  • • Condition the column; stay below its temperature limit
Peak tailing
Likely causes
  • • Active sites in a dirty liner or column inlet
  • • Poorly cut column end or wrong insertion depth
  • • Column contamination
What to do
  • • Replace the liner (deactivated)
  • • Recut the column squarely and reinstall at the correct depth
  • • Trim the front of the column
Retention time shifting
Likely causes
  • • Leak at the inlet or septum
  • • Changed column length after trimming
  • • Wrong carrier gas pressure or flow
  • • Oven temperature not reproducible
What to do
  • • Leak-check and replace the septum
  • • Update column dimensions in the method and adjust flow
  • • Verify flow with a calibrated flow meter
  • • Check oven calibration
Poor area precision
Likely causes
  • • Septum leak or cored septum
  • • Syringe plunger sticking or air bubbles
  • • Liner packed with septum fragments
  • • Split ratio drifting
What to do
  • • Replace the septum
  • • Replace or clean the syringe; adjust draw speed
  • • Replace the liner
  • • Check split vent flow and trap
Split or broad peaks
Likely causes
  • • Solvent and stationary phase poorly matched in splitless mode
  • • Initial oven temperature too high for solvent focusing
  • • Column installed too high in the inlet
What to do
  • • Use a solvent compatible with the phase or a retention gap
  • • Start the oven 10–20 °C below the solvent boiling point
  • • Reinstall at the correct insertion depth
FID noise or spikes
Likely causes
  • • Dirty jet or collector
  • • Incorrect hydrogen/air ratio
  • • Particles from a broken column or septum
What to do
  • • Clean or replace the jet
  • • Reset gas flows to the manufacturer's values
  • • Clean the detector and replace the septum
In LimsPulse · Calibration Management

Keep every GC in control: usage logs, calibration and maintenance schedules, and automatic blocking of overdue instruments.

See how