Spectroscopy
Fourier Transform Infrared (FTIR) Spectrometer
Records the infrared absorption spectrum of a sample, which acts as a molecular fingerprint. The standard technique for identification of APIs and excipients in pharma QC.
- Pharmaceutical
- Polymers & plastics
- Chemicals
- Forensics
- Food
- Academic research

PREVENTIVE MAINTENANCE
Maintenance checklist
Daily / each use
- Check the humidity indicator
- Clean the ATR crystal before and after use
- Inspect the background spectrum for unusual water or contamination bands
Weekly
- Check the energy or interferogram signal against its normal value
- Clean the KBr press die and mortar
Monthly / quarterly
- Replace or regenerate desiccant as indicated
- Run a polystyrene performance check if not done more often
Six-monthly / annual
- Full calibration: wavenumber accuracy, resolution and signal-to-noise
- Preventive maintenance by a trained engineer (source, laser, alignment)
SAFETY
Safety precautions
Before you work on the FTIR
- Do not look into the reference laser beam (a low-power Class 2 or similar laser in most instruments).
- Wear gloves when handling KBr and samples; some APIs are potent.
- Use the hydraulic press with its guard and do not exceed its rated load.
- Clean the ATR crystal with small amounts of solvent away from ignition sources.
- The source runs hot; let it cool before any service access.
In LimsPulse · Calibration Management
Keep every FTIR in control: usage logs, calibration and maintenance schedules, and automatic blocking of overdue instruments.