TROUBLESHOOTING

Common problems, causes and fixes

Weak or no signal
Likely causes
  • • Lamp failed or near end of life
  • • Wrong wavelengths or slits
  • • Shutter closed
  • • PMT voltage too low
What to do
  • • Check lamp and replace
  • • Check method settings
  • • Check shutter status
  • • Increase PMT voltage within the linear range
Signal saturating
Likely causes
  • • Concentration too high
  • • PMT voltage too high
  • • Slits too wide
What to do
  • • Dilute the sample
  • • Lower PMT voltage
  • • Narrow the slits
Calibration curve bends at high concentration
Likely causes
  • • Inner-filter effect
  • • Self-quenching
What to do
  • • Work at lower concentrations
  • • Keep absorbance at λex low
Unexpected peak in blank
Likely causes
  • • Solvent Raman band
  • • Fluorescent impurity in solvent or cuvette
  • • Second-order scatter
What to do
  • • Change λex to move the Raman band
  • • Use fluorescence-grade solvent and clean cuvettes
  • • Use an emission cut-off filter
Readings drift
Likely causes
  • • Lamp not warmed up
  • • Sample temperature changing
  • • Photobleaching
What to do
  • • Allow full warm-up
  • • Use a thermostatted holder
  • • Reduce slit width or exposure time