LimsPulse
Spectroscopy

Fluorescence Spectrophotometer (Spectrofluorometer)

Shines light of one wavelength on a sample and measures the longer-wavelength light the molecules give back (fluorescence). Very sensitive and selective for fluorescent compounds.

  • Pharmaceutical
  • Biotech
  • Clinical
  • Environmental
  • Food & beverage
  • Research
Photo of a Fluorescence Spectrophotometer (Spectrofluorometer)
Photo: VGV27, CC BY-SA 4.0 (cropped), via Wikimedia Commons
CALIBRATION

Fluorimeter calibration procedures

QUALIFICATION

Instrument qualification (DQ · IQ · OQ · PQ)

  1. 01

    DQ — Design qualification

    Defines wavelength range, sensitivity, scan speed, temperature control and software compliance needs.

  2. 02

    IQ — Installation qualification

    Confirms correct installation, components, lamp, software version and documentation.

  3. 03

    OQ — Operational qualification

    Tests excitation and emission wavelength accuracy, sensitivity (water Raman signal-to-noise), repeatability and linearity.

  4. 04

    PQ — Performance qualification

    Shows routine fitness, e.g. a check standard or water Raman intensity tracked over time.

In LimsPulse · Calibration Management

Keep every Fluorimeter in control: usage logs, calibration and maintenance schedules, and automatic blocking of overdue instruments.

See how