Petroleum testing
Flash Point Tester (Pensky-Martens Closed Cup)
Heats a fuel or oil in a closed brass cup while stirring and dips a small flame or electric igniter into the vapour space at set intervals. The lowest temperature at which the vapour flashes is the flash point, a key safety and quality property.
- Petroleum refining
- Fuel terminals and depots
- Lubricant blending
- Marine fuel testing
- Biodiesel production
- Power utilities (transformer oil)
- Independent testing laboratories

TROUBLESHOOTING
Common problems, causes and fixes
Flash point lower than expected
Likely causes
- • Solvent residue left in cup
- • Sample contaminated with a lighter product
- • Sample overfilled or lid wetted
- • Heating rate too high
What to do
- • Clean and dry cup thoroughly
- • Re-sample and check the chain of custody
- • Refill exactly to the mark
- • Check the heating programme and procedure selection
Flash point higher than expected
Likely causes
- • Light ends lost from the sample (warm storage, open container, half-empty bottle)
- • Draught in the test area
- • Ignition source too small or not dipping fully
- • Barometric correction applied with wrong units
What to do
- • Use a fresh, cool, well-filled sample
- • Switch off extraction during the test or use a draught shield
- • Adjust flame size or replace igniter; check shutter mechanism
- • Enter pressure in kPa and recheck the correction
No flash detected although flash seen
Likely causes
- • Flash detector dirty or misaligned
- • Detector sensitivity set wrongly
What to do
- • Clean and reposition the detector
- • Set sensitivity per the manual and verify with a CRM
Sample foams or spatters
Likely causes
- • Water in the sample
- • Heating too fast for a viscous sample
What to do
- • Dry or treat the sample as the method allows
- • Use Procedure B for residual or viscous samples
Flame goes out during the test
Likely causes
- • Gas pressure low
- • Pilot flame not lit
- • Draught
What to do
- • Check gas supply and regulator
- • Relight and check pilot
- • Remove draughts
CRM result outside tolerance
Likely causes
- • Temperature sensor drift
- • Barometer error
- • Worn cup or lid
What to do
- • Calibrate the sensor
- • Check the barometer
- • Replace worn parts and repeat the CRM