Flame Photometer

Measures sodium, potassium and a few other alkali and alkaline earth metals by spraying a solution into a flame and measuring the light the excited atoms emit at their characteristic wavelengths.

  • Environmental and water testing
  • Agriculture and soil testing
  • Fertiliser manufacture
  • Food and beverage
  • Cement and glass (alkali content)
Photo of a Flame Photometer
Photo: A.KRÜSS Optronic GmbH, CC BY-SA 3.0 de (cropped), via Wikimedia Commons
CALIBRATION

Flame photometer calibration procedures

QUALIFICATION

Instrument qualification (DQ · IQ · OQ · PQ)

  1. 01

    Design qualification (DQ)

    Specify elements and ranges required (for example Na and K, 0–100 mg/L), fuel type, filters and data output.

  2. 02

    Installation qualification (IQ)

    Install with a safe fuel supply, regulator, exhaust and compressor; record serial numbers and check for gas leaks.

  3. 03

    Operational qualification (OQ)

    Check zero stability, drift over 15–30 minutes, linearity with Na and K standards and detection limit against the manufacturer's specifications.

  4. 04

    Performance qualification (PQ)

    Analyse a certified water reference material or spiked samples and show recovery and repeatability meet lab limits.