Gel electrophoresis system
A tank with buffer and electrodes, a gel and a DC power supply. Charged molecules such as proteins or DNA move through the gel in the electric field and separate by size, then are stained and imaged.
- Biotechnology
- Biopharmaceutical
- Academic research
- Diagnostics development
- Food (protein profiling)
- Vaccines

Gel electrophoresis calibration procedures
Instrument qualification (DQ · IQ · OQ · PQ)
- 01
DQ
Define gel formats, maximum voltage and current, timer, safety interlocks and imaging needs.
- 02
IQ
Check components, leads, power supply labelling and electrical safety, and record serial numbers.
- 03
OQ
Verify power supply voltage, current and timer accuracy with a calibrated meter and load; check interlocks and leak detection.
- 04
PQ
Run a reference protein or DNA ladder and reference standard; confirm resolution and band pattern meet set criteria.
Keep every Gel electrophoresis in control: usage logs, calibration and maintenance schedules, and automatic blocking of overdue instruments.