Differential Scanning Calorimeter (DSC)

Measures the heat flowing into or out of a small sample compared with an empty reference pan while both are heated or cooled. It shows melting, crystallisation, glass transitions and polymorphic changes.

  • Pharmaceutical
  • Polymers and plastics
  • Food
  • Materials research
  • Chemicals
Photo of a Differential Scanning Calorimeter (DSC)
Photo: Nick Birse, CC BY-SA 4.0 (cropped), via Wikimedia Commons
TROUBLESHOOTING

Common problems, causes and fixes

Noisy or drifting baseline
Likely causes
  • • Contaminated sensor
  • • Unstable purge gas flow
  • • Pan not sitting flat
What to do
  • • Clean the cell by bake-out
  • • Check gas regulator and flow
  • • Use flat, undamaged pans
Indium onset out of limits
Likely causes
  • • Calibration out of date
  • • Wrong heating rate or pan type compared with the calibration
  • • Contaminated indium
What to do
  • • Recalibrate
  • • Calibrate under the same conditions as the method
  • • Use fresh indium
Unexpected sharp exotherm
Likely causes
  • • Decomposition
  • • Pan burst or lid moved
What to do
  • • Check the sample by TGA
  • • Use vented pans and lower end temperature
Broad, unclear melting peak
Likely causes
  • • Large or uneven sample
  • • Heating rate too high
  • • Impure or mixed polymorphic sample
What to do
  • • Use a smaller, flatter sample
  • • Reduce the heating rate
  • • Confirm with XRPD or microscopy
Small endotherm near 100 °C
Likely causes
  • • Water in the sample or condensation in the cell
What to do
  • • Use a pierced lid or dry the sample
  • • Check the purge gas is dry