Atomic Absorption Spectrophotometer (AAS)

Measures one metal at a time by how much light of that element's own wavelength is absorbed by free atoms in a flame or graphite furnace. Simple, robust and still widely used for metals in pharma QC.

  • Pharmaceutical
  • Food & beverage
  • Environmental
  • Mining & metallurgy
  • Agriculture
  • Clinical
Photo of a Atomic Absorption Spectrophotometer (AAS)
Photo: Talos (German Wikipedia), CC BY-SA 3.0 (cropped), via Wikimedia Commons
TROUBLESHOOTING

Common problems, causes and fixes

Low absorbance (poor sensitivity)
Likely causes
  • • Nebuliser partly blocked
  • • Burner not aligned or wrong height
  • • Lamp weak or wrong current
  • • Wrong fuel/oxidant ratio
What to do
  • • Clean or replace the nebuliser
  • • Re-optimise burner position with a standard
  • • Check lamp energy and replace if needed
  • • Re-optimise gas flows
Noisy or drifting signal
Likely causes
  • • Lamp not warmed up
  • • Salt build-up on burner slot
  • • Unstable aspiration rate
What to do
  • • Allow full lamp warm-up
  • • Clean the burner head
  • • Check capillary tubing and nebuliser
Results biased high in a heavy matrix
Likely causes
  • • Background absorption not corrected
  • • Spectral overlap
What to do
  • • Switch on deuterium or Zeeman correction
  • • Use an alternative line or matrix-matched standards
Results biased low for Ca or Mg
Likely causes
  • • Chemical interference from phosphate or silicate
What to do
  • • Add a lanthanum or strontium releasing agent
  • • Use a nitrous oxide–acetylene flame
Flame will not ignite
Likely causes
  • • Low gas pressure
  • • Burner head not seated
  • • Drain trap interlock open
  • • Exhaust interlock
What to do
  • • Check cylinder and regulator
  • • Re-seat burner head
  • • Fill the drain trap
  • • Confirm exhaust is on